Surface damage to cobalt-chrome femoral head prostheses

Author:

Jasty M,Bragdon CR,Lee K,Hanson A,Harris WH

Abstract

Fifty-four cobalt-chrome alloy femoral heads were retrieved at revision surgery or at post-mortem; 23 came from uncemented, eight from hybrid and 23 from cemented hip arthroplasties. The uncemented and hybrid implants had porous coating, metal backing and modular femoral heads; with one exception none of the cemented implants had any of these. Twenty-five of the 31 heads from uncemented and hybrid arthroplasties, and 11 of the 23 heads from cemented arthroplasties showed surface damage involving more than 25% of the surface. Scanning electron microscopy revealed multidirectional fine scratches 1 micron to 10 microns in depth and width which appeared to have been made by fine, hard particles. There was a higher rate of such damage in the uncemented and hybrid arthroplasties than in the cemented implants, suggesting that the abrasive particles were mainly released from the metal, rather than from the cement or polyethylene components of the implants.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery