Author:
Maksimiuk N N,Rebezov M B,Tretyak L N,Varivoda A A,Artyukhova S I,Tolstoguzova T T
Abstract
Abstract
This paper considers and analyzes the results of examining samples for cadmium content by determining them on a Kvant-2AT atomic absorption spectrometer, taking into account the use of the PLP-01M microwave laboratory system and during sample preparation in accordance with GOST 26929 “Raw materials and food products. Sample preparation. Mineralization to determine the content of toxic elements”. The PLP-01M microwave laboratory system uses a fundamentally new method of sample preparation. The decomposition was carried out in a closed system - sealed fluoroplastic vessels under the influence of high temperature, pressure and microwave field. The microwave field in the working chamber of the furnace was created by a special generator-magnetron. Based on the results of the assessment of the operational control of the measurement procedure using samples for cadmium control during the implementation of the PLP-01M microwave laboratory system in the laboratory, the analysis procedure was recognized as satisfactory.
Reference28 articles.
1. Heavy metal pollution in the derwent estuary: history, science and management;Macleod;Regional Studies in Marine Science,2019
2. Determination of inorganic arsenic in white fish using microwave-assisted al-kaline alcoholic sample dissolution and HPLC-ICP-MS;Larsen;Anal. and Bioanal. Chem,2005
3. Heavy metal concentrations in commercially valuable fishes with health hazard inference from Karnaphuliriver, Bangladesh. Human and Ecological Risk Assessment;Ali,2019
4. Heavy metal contamination of river Yamuna, Haryana, India: assessment by metal enrichment factor of the sediments;Kaushik;Journal of Hazardous Materials,2009
5. Food survey: levels and potential health risks of chromium, lead, zinc and copper content in fruits and vegetables consumed in Algeria;Cherfi;Food and Chemical Toxicology,2014
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献