A comparison of physical image quality of two hologic digital mammography systems that utilise linear and 2D anti-scatter grids

Author:

Moore Craig SORCID,Wood Tim JORCID,Saunderson John R,Beavis Andrew WORCID

Abstract

Abstract Objective. Full-field digital mammography (FFDM) systems manufactured by Hologic that utilise either a 2D or linear anti-scatter grid have recently been installed in our clinic. The manufacturer advise that for matched dose, both grids deliver comparable image quality. The aim of this study was to test the manufacturer’s claim using advanced physical image quality metrics and to inform whether the different grids are indeed dose neutral. Approach. Effective detective quantum efficiency (eDQE), effective noise equivalent quanta (eNEQ) and effective dose efficiency (eDE) were measured on a Hologic Dimensions (2D grid) and a Hologic 3Dimensions (linear grid) FFDM system, both calibrated at installation to provide matched threshold contrast, according to the EUREF protocol. eDQE, eNEQ and eDE were calculated and compared using 2, 4, 6 and 7 cm thicknesses of poly (methyl methacrylate) (PMMA) to simulate a clinically appropriate range of breast thicknesses. The beam qualities (target/filter and kilovoltage) chosen were identical between the two systems. Main results. All image quality metrics investigated show that the 2D grid outperforms the linear grid across all spatial frequencies. Furthermore, mean glandular dose (MGD) must be increased by up to 38% on those units that utilise the linear grid if eNEQ is to be matched, although MGD to the standard breast remains within NHSBSP tolerance and below the UK diagnostic reference level. The gradient and shape of each curve was the same irrespective of which grid was used, suggesting that subtle lesions (low frequency information) and micro-calcifications (high frequency information) will be imaged just as efficiently with a linear or 2D grid. Significance. If image quality is to be matched between those units utilising 2D and linear grids, dose must be increased on the latter. This information will be useful to the medical physicist tasked with the optimisation and standardisation of Hologic FFDM units.

Publisher

IOP Publishing

Subject

General Nursing

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Survey of image processing settings used for mammography systems in the United Kingdom: how variable is it?;17th International Workshop on Breast Imaging (IWBI 2024);2024-05-29

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3