Study of the performance change in digital mammography systems depending on the total number of examinations

Author:

Kaya Karaaslan MelikeORCID,Muzoğlu Nedim,Gündoğdu Özcan

Abstract

Abstract Digital mammography equipment should operate at a high-performance level for detecting breast cancer over the lifetime of the systems. This study investigates the performance of the Fujifilm Amulet Innovality full-field digital mammography system in relation to the number of exposures. The performance of twelve systems, from new and up to 17 000 examinations, was compared. The x-ray output, half-value layer (HVL), contrast-detail (C-D) detectability, contrast-to-noise ratio (CNR), mean glandular dose (MGD), signal transfer property (STP), pre-sampled modulation transfer function (MTF), normalized noise power spectra (NNPS), detective quantum efficiency (DQE) tests were performed to determine the performance change. The noise sources were also analyzed in the spatial domain. The x-ray output and HVL values at 28 kV with tungsten anode and rhodium filter were 15 μGy/mAs ± 1.0 (coefficient of variation (cov) = 7.0%) and 0.54 mmAl ± 0.008 (cov = 1.5%), respectively. The average MGDs for 60 mm equivalent breast to reach the achievable image quality level for 0.1 mm and 0.25 mm detail diameters of circular discs on the CDMAM 3.4 phantom image were 1.17 mGy ± 0.13 (cov = 11%) and 1.40 mGy ± 0.09 (cov = 6%), respectively. The average MGD for 53 mm equivalent breast was 1.08 mGy ± 0.14 (cov = 13%) at dose setting Normal in automatic exposure control (AEC) mode. All tested devices demonstrated good linearity with R2  ≥ 0.999 in the STP curves. The average MTF at 5.0 mm−1 spatial frequency was 0.66 ± 0.007 (cov = 1.0%). The average NNPS was 2.38 × 10−6 mm2 ± 2.13 × 10−7 (cov = 9.0%) at 5.0 mm−1. The average DQE value at 5.0 mm−1 was 0.33 ± 0.02 (cov = 5.0%), and the mean peak DQE was 0.66 ± 0.03 (cov = 5.0%). The mean power coefficient (b), determined from the power relationship between linearised standard deviation and detector dose while analyzing noise sources, was 0.49 ± 0.007 (cov = 1.4%). There was no gradual change in the x-ray tube and detector performance by the number of exposures, and taking up to 68 000 images did not decrease the quality of the systems.

Publisher

IOP Publishing

Subject

General Nursing

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Effect of the Anti-scatter Grid on the Detective Quantum Efficiency in Digital Mammography;2023 Medical Technologies Congress (TIPTEKNO);2023-11-10

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3