This study aimed to compare absolute calibration outputs based on the protocols of the International Atomic Energy Agency (IAEA) Technical Report Series (TRS)-398, the American Association of Physicists in Medicine (AAPM) Task Group (TG)-51, and modified calibration approach. The electron beam output calibration followed the IAEA TRS-398 and AAPM TG-51 protocols, both of which required cylindrical chambers and parallel plates. However, the use of cylindrical chambers is not recommended at low energies because of the large fluence-correction factor. TG-51 recommended cross-calibration of the parallel-plate chamber against the cylindrical chamber in a high-energy electron beam. In 2020, an electron beam dosimetry modification was introduced that used a cylindrical ionisation chamber at all energies and updated the data for beam quality conversion factors. This modification provided a lower deviation than that reported in AAPM TG-51. Thus, the modified calibration based on TRS-398 was applied in the present study, which yielded results below the permissible tolerance. The beam calibration at 6, 8, 10, 12, and 15 MeV energies was carried out for two Elekta linear accelerators.. Electron beam dosimetry followed the AAPM TG-51 and TRS-398 protocols, and modified calibration were performed to measure the dose at the maximum depth expressed in dose/monitor units (cGy/MU). Charge-reading measurements were measured using ionisation chambers PTW 30013, IBA CC13, and Exradin A11. The average absorbed dose ratios were 1.004 and 1.009 using the modified calibration and TRS-398 and modified calibration and TG-51, respectively. Therefore, based on IAEA TRS-398, the results were below the tolerance limit (±2%).
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献