Author:
Thabit Hammam Abdurabu,Kabir Norlaili A.,Ismail Abd Khamim,Saleh Muneer Aziz
Abstract
Abstract
The dosimetric properties of synthetic ZnO/Ag/ZnO multilayer film are investigated. The proposed dosimeter was prepared by radio frequency and direct current RF/DC sputtering and irradiated with X-ray doses up to 4 Gy. The properties of thermoluminescence (TL) such as glow curve, dose-response, homogeneity batches, sensitivity, minimum detectable dose (MMD), precision, kinetic parameters (activation energy E, frequency factor S), and percentage depth dose (PDD) were studied. The thin film appeared to have an excellent linear response, and the sensitivity was almost twice the commercial TLD. The readout of the homogeneity and PDD are the same properties of TLD-100. These desirable qualities demonstrated the versatility of this novel synthetic thin film in applications involving radiation detection.
Subject
General Physics and Astronomy
Reference15 articles.
1. Pathological effects of ionizing radiation: endothelial activation and dysfunction;Baselet;Cellular and molecular life sciences,2019
2. Health impacts of low-dose ionizing radiation: current scientific debates and regulatory issues;Vaiserman;Dose-Response,2018
3. Halide lead perovskites for ionizing radiation detection;Wei;Nature communications,2019
4. Radionuclides Concentrations in Some Rock Samples Used as Building Material from Yafea Area in Yemen;El-Kameesy;Arab Journal of Nuclear Sciences and Applications,2016
5. Thermoluminescence studies of CaSO4:Dy,P,Si phosphor under X-ray irradiation;Nair;Ceramics International,2018
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献