Two-time Calibration at Both Ends for Sulfur Measurement by Using High Frequency Induction Infrared Absorption Method

Author:

Wei Jianhuan

Abstract

Abstract High frequency induction heating infrared absorption method is a relative measurement method, and it requires calibrating the analyzer during the measurement process. Usually, one-time calibration on an analyzer is made for high sulfur value by using standard substances. The method proposed by this paper calibrates carbon-sulfur analyzers with a method of two-time calibration at both ends, which eliminates the impact of sulfur blank and thus improves the accuracy and precision of ultra-low sulfur measurement. The method extents the lower limit of measurement of sulfur in high temperature alloy, to 0.00005%, namely, 0.5ppm.

Publisher

IOP Publishing

Subject

General Physics and Astronomy

Reference7 articles.

1. Application of High Frequency Induction Combustion-Infrared Absorption Spectroscopy for Analysis of Carbon or Sulfur in Metal Materials;Liu;Physical Testing and Chemical Analysis Part B:Chemical Analysis,2016

2. Applied Technology with methods for Measuring Sulfur Blank Values by Using Infrared Absorption Method;Tang;Advanced materials Research[J],2014