Author:
Cheng Shikui,Zhang Jie,Zhang Yipo,He Li,Xu Hongbing,Zhu Yuxuan,Zhan Xuwen,Shi Zhongbing,Zhong Wulv
Abstract
Abstract
The development of hard X-ray (HXR) spectrometers with full digital data acquisition for runaway electron studies on the HL-2M tokamak is presented. Consisting of large-sized NaI:Tl detectors and high-performance multichannel analyzers, this system can realize the measurement of HXR with energies ranging from 0.5 to 10 MeV, and spectrum temporal resolution of 1 ms. An average energy resolution of 56 keV at 662 keV has been achieved at counts rates of about 100 kcps. The spectra data is acquired by a digital-based data acquisition system with fast digitization and software signal processing technology. The energy calibrations are performed by using several radioisotope gamma-ray sources, and the detection efficiencies are simulated by MCNP code. First experimental results from HL-2M Ohmic discharges are presented in this paper.
Subject
Mathematical Physics,Instrumentation
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献