Abstract
Engineering Materials Diffractometer (EMD) is included in
Project Phase Two of the China Spallation Neutron Source (CSNS),
which focuses on strain scanning, thermomechanical investigation and
phase transformation analysis under different sample environments,
as well as other novel industrial research for engineering
materials. After a comprehensive analysis and optimization, the
total length of EMD is designed to be 51.5 m where the primary and
secondary flight path are 49.5 m and 2 m, respectively. The
neutron optical components mainly consist of insert guides, curved
guides, bandwidth choppers, straight guides (including both straight
and focusing sections), slits, collimators, detectors, etc. Particularly, an individual 9 m logarithmic spiral focusing guide
to optimize instrument performance was confirmed by Monte Carlo
simulation with the McStas package. Under the target power of
100 kW, for high resolution mode, the resolution Δ d/d is
about 0.25% for 2θ=90∘ and the neutron flux is
4.5× 106 n/cm2/s at the sample position; for high
intensity mode, the neutron flux is 1.4× 107 n/cm2/s
and the resolution is about 0.50%. The high instrument flexibility
could allow and extend a broad of applications in materials science
and engineering fields in the future. This paper will discuss the
details of EMD instrument optical system design especially the
focusing guide and combinations of slits.