Author:
Gavrin V.N.,Ibragimova T.V.,Kozlova J.P.,Tarasov V.A.,Veretenkin E.P.,Zvir A.I.
Abstract
Abstract
Experiments to study non-standard properties of neutrinos using high-intensity artificial neutrino sources require high accuracy in determining their activity. A calorimetric system to determine the activity of a ^51Cr neutrino source in the experiment BEST with an accuracy better than 1% was created. The design of the calorimetric system and the main factors affecting the accuracy of measurements are considered. The results of measurements of the activity of the source made for the BEST experiment are presented. The value of the source activity from calorimetric measurements was 3.41 MCi at the beginning of the experiment.
Subject
Mathematical Physics,Instrumentation
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献