We propose a novel method for exploring the intrinsic
energy resolution of a liquid scintillator (LAB + 2.5 g/L PPO +
3 mg/L bis-MSB) to approximately 1 MeV electrons. With the help of
coincidence detection technology, single-energy electrons of
207Bi were effectively selected. With a careful measurement and
analysis of the energy resolution of a small liquid scintillator
detector, the intrinsic energy resolution of the liquid scintillator
to 976 keV electrons was extracted to be 1.83% ± 0.06%. We
used the wide-angle Compton coincidence (WACC) method to measure the
luminescent nonlinearity of a liquid scintillator and found that it
contributes only weakly to the intrinsic energy resolution of
electrons. Such an unexpected large intrinsic energy resolution may
come from fluctuations in energy transfer processes.