Author:
Zhou Z.,Feng C.,Zhao M.,Wang X.,Li L.,Jiang Z.,Wang Y.,Yuan J.,Cao P.,An Q.
Abstract
Abstract
In-situ radiation measurement and radionuclide
identification in the ocean provide essential information for seabed
geological and marine ecological exploration. In this work, a
small-sized CdZnTe-based gamma spectrometer with high energy
resolution was designed for automatic in-situ radioactivity
measurement in the marine environment down to a depth of
4500 m. Four 10 × 10 × 5 mm3 CdZnTe detectors were
used, and their energy spectra were combined to improve the
detection efficiency. Instead of multi-stage analog circuits,
digital processing methods such as digital filter, trigger, and
energy calculation were implemented in the readout electronics to
improve compactness and reduce the power consumption of the
spectrometer. Monte Carlo simulations were carried out to estimate
the gamma attenuation through the waterproof shell and the marine
efficiency of the spectrometer. Several gamma sources and samples
were used to evaluate the nuclide identification capability of the
spectrometer. The resolution (FWHM) of the spectrometer was 1.8% at
662 keV at room temperature and improved to 1.5% at about
3°C, which is close to the ambient temperature in the
ocean. The spectrometer was tested under a pressure of up to 58 MPa
and was used for marine measurement at a depth of 1418 m in the
South China Sea. The radioactivity of the subsea region is higher
than that of seawater, where thorium daughters, radon daughters, and
^40K were identified in the spectrum result.
Subject
Mathematical Physics,Instrumentation
Cited by
2 articles.