Author:
Deufel Christopher,Dodoo Christopher,Kavanaugh James,Finley Randi,Lang Karen,Sorenson Kasie,Spreiter Sheri,Brooks Jamison,Moseley Douglas,Ahmed Safia K,Haddock Michael G,Ma Daniel,Park Sean S,Petersen Ivy A,Owen Dawn W,Grams Michael P
Abstract
Abstract
Objective. An algorithm was developed for automated positioning of lattice points within volumetric modulated arc lattice radiation therapy (VMAT LRT) planning. These points are strategically placed within the gross tumor volume (GTV) to receive high doses, adhering to specific separation rules from adjacent organs at risk (OARs). The study goals included enhancing planning safety, consistency, and efficiency while emulating human performance. Approach. A Monte Carlo-based algorithm was designed to optimize the number and arrangement of lattice points within the GTV while considering placement constraints and objectives. These constraints encompassed minimum spacing between points, distance from OARs, and longitudinal separation along the z-axis. Additionally, the algorithm included an objective to permit, at the user’s discretion, solutions with more centrally placed lattice points within the GTV. To validate its effectiveness, the automated approach was compared with manually planned treatments for 24 previous patients. Prior to clinical implementation, a failure mode and effects analysis (FMEA) was conducted to identify potential shortcomings. Main results. The automated program successfully met all placement constraints with an average execution time (over 24 plans) of 0.29 ±0.07 min per lattice point. The average lattice point density (# points per 100 c.c. of GTV) was similar for automated (0.725) compared to manual placement (0.704). The dosimetric differences between the automated and manual plans were minimal, with statistically significant differences in certain metrics like minimum dose (1.9% versus 1.4%), D5% (52.8% versus 49.4%), D95% (7.1% versus 6.2%), and Body-GTV V30% (20.7 c.c. versus 19.7 c.c.). Significance. This study underscores the feasibility of employing a straightforward Monte Carlo-based algorithm to automate the creation of spherical target structures for VMAT LRT planning. The automated method yields similar dose metrics, enhances inter-planner consistency for larger targets, and requires fewer resources and less time compared to manual placement. This approach holds promise for standardizing treatment planning in prospective patient trials and facilitating its adoption across centers seeking to implement VMAT LRT techniques.
Reference24 articles.
1. Safety and efficacy of lattice radiotherapy in voluminous non-small cell lung cancer;Amendola;Cureus,2019
2. A current review of spatial fractionation: back to the future?;Billena;Int. J. Radiat. Oncol. Biol. Phys.,2019
3. Evaluation of a commercially-available block for spatially fractionated radiation therapy;Buckey;J. Appl. Clin. Med. Phys.,2010
4. Clinical outcomes of spatially fractionated GRID radiotherapy in the treatment of bulky tumors of the head and neck;Choi;Cureus,2019
5. Spatially fractionated stereotactic body radiation therapy (Lattice) for large tumors;Duriseti;Adv. Radiat. Oncol.,2021