Evaluation of an MRI linac magnetic isocenter walkout with gantry rotation in the presence of angle-specific corrections

Author:

Lotey Rajiv,Latifi Kujtim,Moros Eduardo GORCID,Feygelman Vladimir

Abstract

Abstract Objective. To reduce the magnetic isocenter position variation with gantry rotation on an 0.35 T MRI-guided linac to a practically negligible level. Approach. Central fRequency (CF) offset, eddy current calibration, cross-term calibration, gradient delay, and gradient offsets are tuned for each MR linac installation at every 30° of gantry rotation and stored in a look-up table (LUT). During treatment, the CF is tuned only once in the beginning at an arbitrary gantry angle. After that, imaging paramters are offset based on the stored LUT values for any given gantry angle. Main results. For the same hardware configuration, the implementation of the gantry-angle-specific parameter corrections reduced the total isocenters range of travel in the transverse plane from 1.1 to 0.3 mm and from 0.8 to 0.2 mm in horizontal and vertical directions, respectively. With the longitudinal shift always being negligible (≤0.2 mm), the radius of the sphere encompassing the isocenter locations was reduced from 0.6 to 0.2 mm. Geometric distortion improved as well; in particular, the gantry-angle-averaged maximum longitudinal distortion within a 35 cm diameter sphere was reduced from 1.4 to 0.8 mm. Since the CF is tuned only once during treatment, imaging may resume promptly after the gantry reaches the next target position. Significance. The MRI-guided linear accelerator was conceived primarily as an instrument for precision image-guided therapy. Thus, it is important to keep the treatment and imaging isocentres as close as possible while minimizing the geometric distortion. The described solution reduces the walkout of the imaging isocenter to a fraction of 1 mm, while keeping geometric distortion in a substantial volume below 1 mm. The approach is robust and does not increase the overall procedure time.

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3