Characterization of susceptibility artifacts in magnetic resonance thermometry images during laser interstitial thermal therapy: dimension analysis and temperature error estimation

Author:

De Landro MartinaORCID,Giraudeau Céline,Verde JuanORCID,Ambarki Khalid,Korganbayev SanzharORCID,Wolf AlexeyORCID,Odéen HenrikORCID,Saccomandi PaolaORCID

Abstract

Abstract Objective. Laser interstitial thermal therapy (LITT) is a minimally invasive procedure used to treat a lesion through light irradiation and consequent temperature increase. Magnetic resonance thermometry imaging (MRTI) provides a multidimensional measurement of the temperature inside the target, thus enabling accurate monitoring of the damaged zone during the procedure. In proton resonance frequency shift-based thermometry, artifacts in the images may strongly interfere with the estimated temperature maps. In our work, after noticing the formation of the dipolar-behavior artifact linkable to magnetic susceptibility changes during in vivo LITT, an investigation of susceptibility artifacts in tissue-mimicking phantoms was implemented. Approach. The artifact was characterized: (i) by measuring the area and total volume of error regions and their evolution during the treatment; and (ii) by comparison with temperature reference provided by three temperature sensing needles. Lastly, a strategy to avoid artifacts formation was devised by using the temperature-sensing needles to implement a temperature-controlled LITT. Main results. The artifact appearance was associated with gas bubble formation and with unwanted treatment effects producing magnetic susceptibility changes when 2 W laser power was set. The analysis of the artifact’s dimension demonstrated that in the sagittal plane the dipolar-shape artifact may consistently spread following the temperature trend until reaching a volume 8 times bigger than the ablated one. Also, the artifact shape is quite symmetric with respect to the laser tip. An absolute temperature error showing a negative Gaussian profile in the area of susceptibility artifact with values up to 64.4 °C was estimated. Conversely, a maximum error of 2.8 °C is measured in the area not-affected by artifacts and far from the applicator tip. Finally, by regulating laser power, susceptibility artifacts formation was avoided, and appreciable thermal damage was induced. Significance. These findings may help in improving the MRTI-based guidance of thermal therapies.

Funder

H2020 European Research Council

Publisher

IOP Publishing

Subject

Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3