Use of secondary air kerma from dose area product and fluoroscopy time for preventive effective dose estimation in interventional radiology procedures for staff

Author:

Indovina Luca,Moretti RobertoORCID,Breschi Laura,Cusumano DavideORCID,Meffe Guenda,Placidi Lorenzo,Nardini Matteo,Reale Roberta,Quici Arturo,Placidi Elisa,Stimato Gerardina,Capotosti AmedeoORCID

Abstract

Abstract Introduction: interventional radiology workers are potentially exposed to high levels of ionizing radiation, therefore preventive dose estimation is mandatory for the correct risk classification of staff. Effective dose (ED) is a radiation protection quantity strictly related to the secondary air kerma (K S), using appropriate multiplicative conversion factors (ICRP 106). The aim of this work is to evaluate the accuracy of K S estimated from physically measurable quantities such as dose-area product (DAP) or fluoroscopy time (FT). Methods: radiological units (n = 4) were characterized in terms of primary beam air kerma and DAP-meter response, consequently defining a DAP-meter correction factor (CF) for each unit. K S, scattered from an anthropomorphic phantom and measured by a digital multimeter, was then compared with the value estimated from DAP and FT. Different combinations of tube voltages, field sizes, current and scattering angles were used to simulate the variation of working conditions. Further measurements were performed to estimate the couch transmission factor for different phantom placements on the operational couch, defining a CF as the mean transmission factor. Results: when no CFs were applied, the measured K S showed a median percentage difference of between 33.8% and 115.7% with respect to K S evaluated from DAP, and between −46.3% and 101.8% for K S evaluated from FT. By contrast, when previously defined CFs were applied to the evaluated K S, the median percentage difference between the measured K S and the value evaluated from DAP ranged from between −7.94% and 15.0%, and between −66.2% and 17.2% for that evaluated from FT. Conclusion: when appropriate CF are applied, the preventive ED estimation from the median DAP value seems to be more conservative and easier to obtain with respect to the one obtained from the FT value. Further measurements should be performed with a personal dosimeter during routine activities to assess the proper K S to ED conversion factor.

Publisher

IOP Publishing

Subject

Public Health, Environmental and Occupational Health,Waste Management and Disposal,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3