Author:
Zhang Ning,Hu Qingqing,Wang Qian,Ji Qingchen,Zhao Weijing,Wei Rong,Wang Yuzhu
Abstract
The measurement performance of the atom interferometry absolute gravimeter is strongly affected by the ground vibration noise. We propose a vibration noise evaluation scheme using a Michelson laser interferometer constructed by the intrinsic Raman laser of the atomic gravimeter. Theoretical analysis shows that the vibration phase measurement accuracy is better than 4 mrad, which corresponds to about 10−10
g accuracy for a single shot gravity measurement. Compared with the commercial seismometer or accelerometer, this method is a simple, low cost, direct, and fully synchronized measurement of the vibration phase which should benefit the development of the atomic gravimeter. On the other side, limited by equivalence principle, the result of the laser interferometer is not absolute but relative vibration measurement. Triangular cap method could be used to evaluation the noise contribution of vibration, which is a different method from others and should benefit the development of the atomic gravimeter.
Subject
General Physics and Astronomy
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献