Author:
Kasbekar A V,Scoffings D J,Kenway B,Cross J,Donnelly N,Lloyd S W K,Moffat D,Axon P R
Abstract
AbstractObjectives:We evaluated use of the periodically rotated overlapping parallel lines with enhanced reconstruction diffusion-weighted imaging sequence, compared with conventional echo planar magnetic resonance imaging, in the detection of middle-ear cholesteatoma.Material and methods:Sixteen patients awaiting second-stage combined approach tympanoplasty and three patients awaiting first-stage combined approach tympanoplasty underwent magnetic resonance imaging with both (1) the periodically rotated overlapping parallel lines with enhanced reconstruction sequence (i.e. non echo planar imaging) and (2) the array spatial sensitivity encoding technique sequence (i.e. echo planar imaging). Two neuroradiologists independently evaluated the images produced by both sequences. Radiology findings were correlated with surgical findings.Results and analysis:Seven cholesteatomas were found at surgery. Neither of the assessed imaging sequences were able to detect cholesteatoma of less than 4 mm. Rates for sensitivity, specificity, and positive and negative predictive values are presented.Conclusion:Decisions on whether or not to operate for cholesteatoma cannot be made based on the two imaging sequences assessed, as evaluated in this study. Other contributing factors are discussed, such as the radiological learning curve and technical limitations of the magnetic resonance imaging equipment.
Publisher
Cambridge University Press (CUP)
Subject
Otorhinolaryngology,General Medicine
Cited by
29 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献