Author:
Pína L.,Fiedorowicz H.,Koshevoi M. O.,Rupasov A. A.,Rus B.,Shikanov A. S.,Svoboda V.
Abstract
A program is under way to develop methods and instrumentation based on charge-coupled device (CCD) sensors for hot plasma diagnostics. We have developed a new X-ray spectrometer in which a freestanding X-ray transmission grating is coupled to a CCD linear array detector with electronic digitized readout replacing film and its wet processing. This instrument measures time-integrated pulsed X-ray spectra with moderate spectral resolution (δλ ≤ 0.6 nm) over a broad spectral range (0.3–2 keV) with high sensitivity, linearity, and large dynamic range. The performance of the device was tested using laser plasma as the X-ray source.
Subject
Electrical and Electronic Engineering,Condensed Matter Physics,Atomic and Molecular Physics, and Optics