Abstract
Radiocarbon dating of many types of samples has continued at the La Jolla Radiocarbon Laboratory since the last date list was published (Bien and Pandolfi, 1972). Since the 1969 relocation of the laboratory to its present location on Mt Soledad, at least 5 low-level14C systems have generally been in operation. The laboratory has continued to employ acetylene gas proportional counting. Samples were usually measured for 2 to 3 days in each of 2 different detectors. The properties of the 5 counters used in the measurements of samples are as follows (background and net standard counts per minute, respectively, with acetylene of 900mmHg pressure are given in parentheses): 1) Bern counter: Oeschger-type counter-guard ring unit with 1.7L sensitive center volume (2.1, 21.5), 2) MBLE counter: Oeschger-type counter-guard ring unit with 1.6L center volume (1.3, 17.4), 3) 1L counter with separate Q-gas flow guard ring (4.2, 12.9), 4) 0.4L counter with separate guard ring (1.2, 4.2), and 0.1L counter with separate guard ring (0.7, 1.2).
Publisher
Cambridge University Press (CUP)
Subject
General Earth and Planetary Sciences,Archeology
Reference33 articles.
1. Deformation Lamellae Parallel to (101̅3) and (0001) in Quartz of the Coeur d'Alene District, Idaho
2. Historical Dates for Neolithic Sites of Southeast Europe
3. Obre, Yugoslavia–two neolithic sites;Gimbutas;Archaeology,1970
4. Otvos E G Jr 1973, Geology of the Mississippi-Alabama coastal area and nearshore zone: Guidebook, New Orleans Geol Soc, 67 p.
Cited by
26 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献