Author:
Puchades Vicente,Serna Alfredo,Mata-Colodro Fernando,Ramos-Amores Davis,Casal Emilio,Alcaraz Miguel
Abstract
AbstractPurposeThe objective of this work was to investigate the accuracy of AAA dose calculation algorithm for RapidArc volumetric modulated technique (VMAT) in the presence of anatomical heterogeneities in the pelvic region.Material and methodsAn anthropomorphic phantom was used to simulate a prostate case, delineating planning target volumes (PTVs) and organs at risk. VMAT plans were optimised in eclipse (v10·0) treatment planning system (TPS). The dose distributions were calculated by the AAA dose calculation algorithm. A total of 49 thermoluminiscent dosimeters were inserted into the anthropomorphic phantom and dose measurements were compared with the predicted TPS doses.ResultsThe average dose variation was −1·5% for planning target volume corresponding to the prostate and −0·3% for planning target volume corresponding to the pelvic nodes, −0·2% for the rectum, +2·4% for the bladder, −2·0% for the femoral heads and +1·0% for the intestinal package.ConclusionAAA is a reliable dose calculation for the treatment with VMAT in the anatomy of the pelvis.
Publisher
Cambridge University Press (CUP)
Subject
Oncology,Radiology, Nuclear Medicine and imaging
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献