Tracked 3D ultrasound and deep neural network-based thyroid segmentation reduce interobserver variability in thyroid volumetry

Author:

Krönke Markus,Eilers ChristineORCID,Dimova Desislava,Köhler Melanie,Buschner Gabriel,Schweiger Lilit,Konstantinidou LemoniaORCID,Makowski Marcus,Nagarajah James,Navab Nassir,Weber Wolfgang,Wendler ThomasORCID

Abstract

Thyroid volumetry is crucial in the diagnosis, treatment, and monitoring of thyroid diseases. However, conventional thyroid volumetry with 2D ultrasound is highly operator-dependent. This study compares 2D and tracked 3D ultrasound with an automatic thyroid segmentation based on a deep neural network regarding inter- and intraobserver variability, time, and accuracy. Volume reference was MRI. 28 healthy volunteers (24—50 a) were scanned with 2D and 3D ultrasound (and by MRI) by three physicians (MD 1, 2, 3) with different experience levels (6, 4, and 1 a). In the 2D scans, the thyroid lobe volumes were calculated with the ellipsoid formula. A convolutional deep neural network (CNN) automatically segmented the 3D thyroid lobes. 26, 6, and 6 random lobe scans were used for training, validation, and testing, respectively. On MRI (T1 VIBE sequence) the thyroid was manually segmented by an experienced MD. MRI thyroid volumes ranged from 2.8 to 16.7ml (mean 7.4, SD 3.05). The CNN was trained to obtain an average Dice score of 0.94. The interobserver variability comparing two MDs showed mean differences for 2D and 3D respectively of 0.58 to 0.52ml (MD1 vs. 2), −1.33 to −0.17ml (MD1 vs. 3) and −1.89 to −0.70ml (MD2 vs. 3). Paired samples t-tests showed significant differences for 2D (p = .140, p = .002 and p = .002) and none for 3D (p = .176, p = .722 and p = .057). Intraobsever variability was similar for 2D and 3D ultrasound. Comparison of ultrasound volumes and MRI volumes showed a significant difference for the 2D volumetry of all MDs (p = .002, p = .009, p <.001), and no significant difference for 3D ultrasound (p = .292, p = .686, p = 0.091). Acquisition time was significantly shorter for 3D ultrasound. Tracked 3D ultrasound combined with a CNN segmentation significantly reduces interobserver variability in thyroid volumetry and increases the accuracy of the measurements with shorter acquisition times.

Funder

H2020 LEIT Information and Communication Technologies

Publisher

Public Library of Science (PLoS)

Subject

Multidisciplinary

Cited by 19 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Cross-view discrepancy-dependency network for volumetric medical image segmentation;Medical Image Analysis;2025-01

2. Deep learning-based segmentation of 3D ultrasound images of the thyroid;WFUMB Ultrasound Open;2024-12

3. Volumetric Ultrasound Imaging for the Whole Soft Tissue: Toward Enhanced Thyroid Disease Examination;Ultrasound in Medicine & Biology;2024-09

4. Ultrasound segmentation analysis via distinct and completed anatomical borders;International Journal of Computer Assisted Radiology and Surgery;2024-05-25

5. NeSM: A NeRF-Based 3D Segmentation Methodfor Ultrasound Images;2024 10th International Conference on Applied System Innovation (ICASI);2024-04-17