Abstract
Background
A 12-lead electrocardiogram (ECG) and Holter ECG have been established as gold standards for detection of arrhythmias. Recently, wearable ECG monitoring devices have been available. Our purpose of the present study was to investigate whether a novel wearable electrode embedded in an undershirt is useful for ECG monitoring and detection of arrhythmias.
Methods
We studied 31 consecutive hospitalized patients who underwent catheter ablation of tachyarrhythmias. Patients equipped a wearable electrode and a lead CM5 of Holter ECG simultaneously, and total heart beats, maximum heart rate (HR), mean HR, minimum HR, detections of arrhythmias, such as atrial fibrillation, non-sustained ventricular tachycardia, and premature ventricular contractions (Lown’s grade >II), were compared between the two methods using a Holter ECG analysis software.
Results
Median recording time of ECG by wearable electrodes was 12.6 hours. Strong correlations between the two methods were observed in total heart beats (R = 0.999, P <0.001), maximum HR (R = 0.997, P <0.001), mean HR (R = 0.999, P <0.001), minimum HR (R = 0.989, P <0.001) and QRS duration (R = 0.900, P <0.001). Bland-Altman analysis showed excellent concordance between each parameter measured by two methods. In addition, the detection of atrial fibrillation (nine events), non-sustained ventricular tachycardia (two events), and premature ventricular contractions of Lown’s grade >II (five events) were concordant in two methods. In addition, there were no significant difference in parameters of time-domain and frequency-domain analyses of heart rate variability between the two methods.
Conclusions
The usefulness of a novel electrode embedded in an undershirt is equivalent to that of a Holter ECG in monitoring the ECG and detection of arrythmias.
Publisher
Public Library of Science (PLoS)
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献