Affiliation:
1. University of Sarajevo, Faculty of Mechanical Engineering, Vilsonovo šetalište 9, 71000 Sarajevo, Bosnia and Herzegovina
Abstract
In this research, an analysis of the mechanical behaviour for the Orthofix external fixation device under the impact of torque was performed. Research considers application of the Orthofix device on the tibia bone for the case of unstable fracture. 3D (Three Dimensional) model of the Orthofix device was created in the CATIA (Computer Aided Three-Dimensional Interactive Application) software, based on the real device construction. Structural analysis was used to monitor and analyse the stress magnitudes on the specific areas of the fixation device and fracture. With usage of the interfragmentary displacement data for the bone fragments, degrees of stiffness are introduced for the fracture and fixation device. Obtained results are used to specify the mechanical behaviour of the Orthofix fixation device.
Subject
General Materials Science
Cited by
3 articles.
订阅此论文施引文献