Investigation of the effects of pneumatic tube transport system on routine biochemistry, hematology, and coagulation tests in Ankara City Hospital
Author:
Yurt Emine F.1, Akbiyik Filiz2, Bicer Cemile34
Affiliation:
1. Medical Biochemistry , Adilcevaz Oncology Hospital , Bitlis , Turkey 2. Medical Biochemistry , Siemens Healthineers, Ankara City Hospital , Ankara , Turkey 3. Medical Biochemistry , Faculty of Medicine, Ankara Yildirim Beyazit University , Ankara , Turkey 4. Medical Biochemistry , Ankara City Hospital , Ankara , Turkey
Abstract
Abstract
Objectives
Academics are far from a consensus regarding the effects of pneumatic tube system (PTS) delivery on sample integrity and laboratory test results. As for the reasons for conflicting opinions, each PTS is uniquely designed, sample tubes and patient characteristics differ among studies. This study aims to validate the PTS utilized in Ankara City Hospital for routine chemistry, coagulation, and hematology tests by comparing samples delivered via PTS and porter.
Methods
The study comprises 50 healthy volunteers. Blood samples were drawn into three biochemistry, two coagulation, and two hemogram tubes from each participant. Each of the duplicate samples was transferred to the emergency laboratory via Swiss log PTS (aka PTS-immediately) or by a porter. The last of the biochemistry tubes were delivered via the PTS, upon completion of coagulation of the blood (aka PTS-after). The results of the analysis in these groups were compared with multiple statistical analyses.
Results
The study did not reveal any correlation between the PTS and serum hemolysis index. There were statistically significant differences in several biochemistry tests. However, none of them reached the clinical significance threshold. Basophil and large unidentified cell (LUC) tests had poor correlations (r=0.47 and r=0.60; respectively) and reached clinical significance threshold (the average percentages of bias, 10.2%, and 15.4%, respectively). The remainder of the hematology and coagulation parameters did not reach clinical significance level either.
Conclusions
The modern PTS validated in this study is safe for sample transportation for routine chemistry, coagulation, and hematology tests frequently requested in healthy individuals except for basophil and LUC.
Funder
Siemens Healthineers
Publisher
Walter de Gruyter GmbH
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Reference32 articles.
1. Fernandes, CM, Worster, A, Eva, K, Hill, S, McCallum, C. Pneumatic tube delivery system for blood samples reduces turnaround times without affecting sample quality. J Emerg Nurs 2006;32:139–43. https://doi.org/10.1016/j.jen.2005.11.013. 2. Pupek, A, Matthewson, B, Whitman, E, Fullarton, R, Chen, Y. Comparison of pneumatic tube system with manual transport for routine chemistry, hematology, coagulation and blood gas tests. Clin Chem Lab Med 2017;55:1537–44. https://doi.org/10.1515/cclm-2016-1157. 3. Streichert, T, Otto, B, Schnabel, C, Nordholt, G, Haddad, M, Maric, M, et al.. Determination of hemolysis thresholds by the use of data loggers in pneumatic tube systems. Clin Chem 2011;57:1390–7. https://doi.org/10.1373/clinchem.2011.167932. 4. Gomez-Rioja, R, Fernandez-Calle, P, Alcaide, MJ, Madero, R, Oliver, P, Iturzaeta, JM, et al.. Interindividual variability of hemolysis in plasma samples during pneumatic tube system transport. Clin Chem Lab Med 2013;51:e231–3. https://doi.org/10.1515/cclm-2013-0171. 5. Cui, M, Jing, R, Wang, H. Changes of serum lactate dehydrogenase and potassium levels produced by a pneumatic tube system. Lab Med 2009;40:728–31. https://doi.org/10.1309/lmznz2y1p3krdqyl.