Quantitative protein mass-spectrometry requires a standardized pre-analytical phase
Author:
Smit Nico P.M.1ORCID, Romijn Fred P.H.T.M.1, van Ham Vanessa J.J.2, Reijnders Esther1, Cobbaert Christa M.1ORCID, Ruhaak L. Renee1ORCID
Affiliation:
1. Department of Clinical Chemistry and Laboratory Medicine , Leiden University Medical Center , Leiden , The Netherlands 2. Biobank Facility, Leiden University Medical Center , Leiden , The Netherlands
Abstract
Abstract
Objectives
Quantitative protein mass-spectrometry (QPMS) in blood depends on tryptic digestion of proteins and subsequent measurement of representing peptides. Whether serum and plasma can be used interchangeably and whether in-vitro anticoagulants affect the recovery is unknown. In our laboratory serum samples are the preferred matrix for QPMS measurement of multiple apolipoproteins. In this study, we investigated the effect of different matrices on apolipoprotein quantification by mass spectrometry.
Methods
Blood samples were collected from 44 healthy donors in Beckton Dickinson blood tubes simultaneously for serum (with/without gel) and plasma (heparin, citrate or EDTA). Nine apolipoproteins were quantified according to standard operating procedure using value-assigned native serum calibrators for quantitation. Tryptic digestion kinetics were investigated in the different matrices by following formation of peptides for each apolipoprotein in time, up to 22 h.
Results
In citrate plasma recovery of apolipoproteins showed an overall reduction with a bias of −14.6%. For heparin plasma only −0.3% bias was found compared to serum, whereas for EDTA-plasma reduction was more pronounced (−5.3% bias) and variable with >14% reduction for peptides of apoA-I, A-II and C-III. Digestion kinetics revealed that especially slow forming peptides showed reduced formation in EDTA-plasma.
Conclusions
Plasma anticoagulants affect QPMS test results. Heparin plasma showed comparable results to serum. Reduced concentrations in citrate plasma can be explained by dilution, whereas reduced recovery in EDTA-plasma is dependent on altered proteolytic digestion efficiency. The results highlight the importance of a standardized pre-analytical phase for accurate QPMS applications in clinical chemistry.
Publisher
Walter de Gruyter GmbH
Subject
Biochemistry (medical),Clinical Biochemistry,General Medicine
Reference30 articles.
1. Smit, NPM, Ruhaak, LR, Romijn, F, Pieterse, MM, van der Burgt, YEM, Cobbaert, CM. The time has come for quantitative protein mass spectrometry tests that target unmet clinical needs. J Am Soc Mass Spectrom 2021;32:636–47. https://doi.org/10.1021/jasms.0c00379. 2. Cobbaert, CM, Althaus, H, Begcevic Brkovic, I, Ceglarek, U, Coassin, S, Delatour, V, et al.. Towards an SI-traceable reference measurement system for seven serum apolipoproteins using bottom-up quantitative proteomics: conceptual approach enabled by cross-disciplinary/cross-sector collaboration. Clin Chem 2021;67:478–89. https://doi.org/10.1093/clinchem/hvaa239. 3. Ruhaak, LR, Romijn, F, Smit, NPM, van der Laarse, A, Pieterse, MM, de Maat, MPM, et al.. Detecting molecular forms of antithrombin by LC-MRM-MS: defining the measurands. Clin Chem Lab Med 2018;56:1704–14. https://doi.org/10.1515/cclm-2017-1111. 4. van den Broek, I, Romijn, FP, Nouta, J, van der Laarse, A, Drijfhout, JW, Smit, NP, et al.. Automated multiplex LC-MS/MS assay for quantifying serum apolipoproteins A-I, B, C-I, C-II, C-III, and E with qualitative apolipoprotein E phenotyping. Clin Chem 2016;62:188–97. https://doi.org/10.1373/clinchem.2015.246702. 5. van Duijl, TT, Ruhaak, LR, Smit, NPM, Pieterse, MM, Romijn, F, Dolezal, N, et al.. Development and provisional validation of a multiplex LC-MRM-MS test for timely kidney injury detection in urine. J Proteome Res 2021;20:5304–14. https://doi.org/10.1021/acs.jproteome.1c00532.