The aim of this work is to characterize the ferrous sulfate-benzoic acid-xylenol orange (FBX) aqueous chemical dosimeter developed at our laboratory, prepared using ultra pure water, by measuring the absorption spectrum, dose response curve, precision and accuracy, energy and dose rate dependency and stability of response. The FBX readings were evaluated by using an accurate spectrophotometer. Experimental data were obtained using various nominal energies 6 MV, 18 MV, 12 MeV, and 15 MeV, including the 60Co γ-rays beam. The calibration of the dosimeters was performed using the ionization chamber as a reference dosimeter. The results show that the FBX dosimeter has a good precision of about 0.2%, no significant energy, dose rate dependence and a linear dose-response relationship in the 1-5 Gy range.
Reference16 articles.
1. [1] Gupta BL. Low-level dosimetric studies with the FeSO4–benzoic acid–xylenol orange system. In: Dosimetry in agriculture, industry, biology and medicine. Proceedings of a symposium organized by the IAEA [International Atomic Energy Agency] 17021 April 1972; Vienna, Austria. Vienna: IAEA; 1973: 4210432.
2. [2] Gupta BL, Gomathy KR. Consistency of ferrous sulfate-benzoic acid-xylenol orange dosimeter. Int J Appl Radiat Isot. 1974;25(11–12):509-513.
3. [3] Gupta BL, Bhat RM, Gomathy KR Suseela B. Radiation chemistry of the ferrous sulphate-benzoic acid-xylenol orange dosimeter. Radiat Res. 1978;75(2):269-277.10.2307/3574902
4. [4] Gupta BL, Kini UR, Bhat RM, Madhvanath U. Use of the FBX dosimeter for the calibration of cobalt-60 and high energy teletherapy machines. Phys Med Biol. 1982;27(2):235-245.10.1088/0031-9155/27/2/005
5. [5] Gupta BL, Narayan GR, Bhat RM, Kini UR, et al. Use of the FBX dosimeter for the quality assurance of 60Co and high-energy teletherapy machines by mail. Phys Med Biol. 1992;37(11):2095-2102.10.1088/0031-9155/37/11/006
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献