AbstractThe 90 ° - bend represents the most common structural element in piping systems and can have a significant impact on the measuring accuracy of flow meters installed downstream. Within the type-approval procedure of water meters, its impact is emulated by means of a segmental orifice plate with a segment area of 7 % (SOP7). In research and development, coverage rates of 33 % (SOP33) or more are sometimes used as an alternative. The purpose of this study is to evaluate the comparability of the flow conditions and their influence on the measurement deviation of flow meters, which is the basic requirement for using SOPs as a substitute for the 90°- bend. We present laser Doppler measurements downstream of an SOP33 and a 90°- bend and describe the flow development in a distance range from 2 to 30 diameters. Besides a quantitative comparison with performance indicators, the measurements are used to model the response of an ultrasonic and electromagnetic flow meter, including recent investigations of an SOP7. The results demonstrate the consistently poor agreement between the SOP7 and the 90°- bend, whereas the SOP33 provides similar flow conditions starting at a distance of 10 diameters. Further studies are necessary to develop a disturbance generator emulating the near-field range.