ISCEV standard pattern reversal VEP development: paediatric reference limits from 649 healthy subjects

Author:

Thompson Dorothy A.ORCID,Mikó-Baráth EszterORCID,Hardy Sharon E.,Jandó GáborORCID,Shaw Martin,Hamilton RuthORCID

Abstract

Abstract Purpose To establish the extent of agreement for ISCEV standard reference pattern reversal VEPs (prVEPs) acquired at three European centres, to determine any effect of sex, and to establish reference intervals from birth to adolescence. Methods PrVEPs were recorded from healthy reference infants and children, aged 2 weeks to 16 years, from three centres using closely matched but non-identical protocols. Amplitudes and peak times were modelled with orthogonal quadratic and sigmoidal curves, respectively, and two-sided limits, 2.5th and 97.5th centiles, estimated using nonlinear quantile Bayesian regression. Data were compared by centre and by sex using median quantile confidence intervals. The ‘critical age’, i.e. age at which P100 peak time ceased to shorten, was calculated. Results Data from the three centres were adequately comparable. Sex differences were not clinically meaningful. The pooled data showed rapid drops in P100 peak time which stabilised by 27 and by 34 weeks for large and small check widths, respectively. Post-critical-age reference limits were 87–115 ms and 96–131 ms for large and small check widths, respectively. Amplitudes varied markedly and reference limits for all ages were 5–57 μV and 3.5–56 μV for large and small check widths, respectively. Conclusions PrVEP reference data could be combined despite some methodology differences within the tolerances of the ISCEV VEP Standard, supporting the clinical benefit of ISCEV Standards. Comparison with historical data is hampered by lack of minimum reporting guidelines. The reference data presented here could be validated or transformed for use elsewhere.

Publisher

Springer Science and Business Media LLC

Subject

Physiology (medical),Sensory Systems,Ophthalmology

Reference53 articles.

1. Odom JV, Bach M, Brigell M et al (2016) ISCEV standard for clinical visual evoked potentials: (2016 update). Doc Ophthalmol 133:1–9. https://doi.org/10.1007/s10633-016-9553-y

2. Horn PS, Pesce AJ (2005) Reference intervals: a user’s guide. AACC Press, Washington, DC

3. Davis CQ, Hamilton R (2021) Reference ranges for clinical electrophysiology of vision. Doc Ophthalmol. https://doi.org/10.1007/s10633-021-09831-1

4. Horowitz GL, Clinical and Laboratory Standards Institute (CLSI) (2010) Defining, establishing, and verifying reference intervals in the clinical laboratory; approved guideline–third edition. CLSI document EP28-A3c. Clinical and laboratory standards institute, Wayne, PA

5. Reed A, Henry R, Mason W (1971) Influence of statistical method used on resulting estimate of normal range. Clin Chem 17:275–284