Abstract
Abstract
Purpose
In the event that any ocular parameter involved in the calculation of intraocular lens power could not be properly measured in one eye, it is important to know whether clinically relevant differences between both eyes can be expected. The aim of this work is to evaluate the symmetry of interocular biometric parameters.
Methods
This was a prospective, cross-sectional study involving 4090 subjects. Patients underwent consecutive swept-source optical biometry performed with an IOLMaster 700 (Carl Zeiss Meditec AG, Jena, Germany). The biometric parameters that were evaluated were: axial length (AL), mean anterior curvature (Rm), anterior chamber depth (ACD), crystalline lens thickness (LT), central corneal thickness (CCT) and white-to-white (WTW). The Chang–Waring chord distance (CWC-D) and the Chang–Waring chord angle (CWC-A) were also evaluated.
Results
There is an excellent correlation between both eyes for almost all the biometric parameters under study, with the exception of the CWC. Agreement for AL was better for eyes shorter than 24 mm. The linearity of the OD-vs-OS relationship can be correctly assumed for all parameters (Cusum test: p > 0.05 in all cases).
Conclusion
There are no clinically significant interocular differences for the biometric parameters under study, although for all of them, except the LT, statistically significant differences did arise. In the case of AL, moderate differences can be expected in eyes larger than 24 mm.
Funder
Universidad Complutense de Madrid
Publisher
Springer Science and Business Media LLC
Cited by
7 articles.