Author:
Berisha Sabina,Živković Igor,Kotnik Jože,Mlakar Tanja Ljubič,Horvat Milena
Abstract
AbstractThis work aimed to assess the efficiency of a portable total mercury (Hg) analyzer (Lumex RA-915 +) that employs direct thermal decomposition atomic absorption spectrometry (AAS) to liberate Hg from a sample matrix as well as validate the quantification method of total Hg in different materials that are used in cement production. Regarding the solid matrix of the samples, the approaches included matrix-matched certified reference materials (CRM) calibration and adding a standard calibration solution to the solid samples. While both calibration approaches were suitable for the instrumentation employed, the uncertainty of the Hg mass fraction in a CRM and the need to include several matrix CRMs that differed in analyte content added to the complexity of the case. Spiked calibration is an acceptable alternative to calibration because it provides a lower expanded uncertainty at 15 %, 15.5 %, and 21.7 %, for the highest (> 200 ng), middle (50–150 ng), and lowest (< 50 ng) concentration ranges, respectively, compared to matrix CRM calibration. The obtained expanded uncertainty, the minimal requirements for sample preparation, easy and straightforward calibration and measurement, and high sample output indicate that the quantification of total Hg with thermal decomposition coupled with AAS is suitable for materials that are used in cement production.
Publisher
Springer Science and Business Media LLC
Subject
Safety, Risk, Reliability and Quality,Instrumentation,General Chemical Engineering,General Chemistry
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献