Abstract
Abstract
Introduction
Gastrointestinal anastomoses with classical sutures and/or metal staples have resulted in significant bleeding and leak rates. This multi-site study evaluated the feasibility, safety, and preliminary effectiveness of a novel linear magnetic compression anastomosis device, the Magnet System (MS), to form a side-to-side duodeno-ileostomy (DI) diversion for weight loss and type 2 diabetes (T2D) resolution.
Methods
In patients with class II and III obesity (body mass index [BMI, kg/m2] ≥ 35.0– ≤ 50.0 with/without T2D [HbA1C > 6.5%]), two linear MS magnets were delivered endoscopically to the duodenum and ileum with laparoscopic assistance and aligned, initiating DI; sleeve gastrectomy (SG) was added. There were no bowel incisions or retained sutures/staples. Fused magnets were expelled naturally. Adverse events (AEs) were graded by Clavien-Dindo Classification (CDC).
Results
Between November 22, 2021 and July 18, 2022, 24 patients (83.3% female, mean ± SEM weight 121.9 ± 3.3 kg, BMI 44.4 ± 0.8) in three centers underwent magnetic DI. Magnets were expelled at a median 48.5 days. Respective mean BMI, total weight loss, and excess weight loss at 6 months (n = 24): 32.0 ± 0.8, 28.1 ± 1.0%, and 66.2 ± 3.4%; at 12 months (n = 5), 29.3 ± 1.5, 34.0 ± 1.4%, and 80.2 ± 6.6%. Group mean respective mean HbA1C and glucose levels dropped to 1.1 ± 0.4% and 24.8 ± 6.6 mg/dL (6 months); 2.0 ± 1.1% and 53.8 ± 6.3 mg/dL (12 months). There were 0 device-related AEs, 3 procedure-related serious AEs. No anastomotic bleeding, leakage, stricture, or mortality.
Conclusion
In a multi-center study, side-to-side Magnet System duodeno-ileostomy with SG in adults with class III obesity appeared feasible, safe, and effective for weight loss and T2D resolution in the short term.
Publisher
Springer Science and Business Media LLC
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献