Abstract
Abstract
Purpose
The current cut-offs for the diagnosis of adrenal insufficiency (AI) have been established using outdated immunoassays. We compared the cortisol concentrations measured with Roche Cortisol I (R1), the newly available Roche Cortisol II (R2), and liquid chromatography tandem mass spectrometry (LC-MS/MS), the gold standard procedure to measure steroids in patients undergoing the corticotropin (ACTH) test.
Methods
We enrolled 30 patients (age 47 ± 21 years) referred to undergo the ACTH test (1 or 250 μg). Cortisol was measured at 0, 30, and 60 min after stimulation with R1, R2, and LC-MS/MS. AI was diagnosed for R1-stimulated peak cortisol concentrations < 500 nmol/L.
Results
Mean cortisol concentrations measured with R2 and LC-MS/MS were comparable, while mean cortisol concentrations measured by R1 were higher than those of both R2 and LC-MS/MS (respectively, basal 411 ± 177, 287 ± 119, and 295 ± 119 nmol/L; at 30 min, 704 ± 204, 480 ± 132, and 500 ± 132 nmol/L; at 60 min, 737 ± 301, 502 ± 196, and 519 ± 201 nmol/L, p ≤ 0.01 for R1 vs. both R2 and LC-MS/MS at each point). Considering the 500 nmol/L cortisol peak cut-off, AI was diagnosed in 5/30 patients using R1 and in 12/30 using R2 (+ 140%). Based on the correlation between R1 and R2, the threshold of 500 nmol/L became 351 nmol/L (12.7 μg/dL) when cortisol was measured with R2, and 368 nmol/L (13.3 μg/dL) with LC-MS/MS.
Conclusions
The use of more specific cortisol assays results in lower cortisol concentrations. This could lead to misdiagnosis and overtreatment when assessing AI with the ACTH test if a different cut-off for cortisol peak is not adopted.
Funder
Ricerca Corrente Funds from the Italian Ministry of Health to Fondazione IRCCS Ca' Granda Ospedale Maggiore Policlinico
Publisher
Springer Science and Business Media LLC
Subject
General Medicine,Endocrinology, Diabetes and Metabolism