Abstract
Abstract
Purpose
This study aimed to evaluate the effectiveness and safety of the NeVaTM stent retriever as first- and second-line device for mechanical thrombectomy in acute ischemic stroke.
Methods
In this retrospective single-center study, all consecutive patients that underwent mechanical thrombectomy with NeVaTM stent retriever as first- or second-line device due to intracranial vessel occlusion with acute ischemic stroke between March and November 2022 were included.
Results
Thirty-nine patients (m=18, f=21) with a mean age of 69.9 ± 13.3 years were treated with the NeVaTM stent retriever. NeVaTM stent retriever was used as first-line device in 24 (61.5%) of patients and in 15 (38.5%) as second-line device. First-pass rate (≥mTICI 2c) of NeVaTM stent retriever was both 66.7% when used as first- or second-line device. Final recanalization rate including rescue strategies was 92.3% for ≥mTICI2c and 94.9% for ≥mTICI2b. No device-related minor or major adverse events were observed. A hemorrhage was detected in 33.3% of patients at 24h post-thrombectomy dual-energy CT, of which none was classified as symptomatic intracerebral hemorrhage. NIHSS and mRS improved significantly at discharge compared to admission (p<0.05).
Conclusion
The NeVaTM stent retriever has a high effectivity and good safety profile as first- and second-line device for mechanical thrombectomy in acute ischemic stroke.
Funder
Westfälische Wilhelms-Universität Münster
Publisher
Springer Science and Business Media LLC
Subject
Cardiology and Cardiovascular Medicine,Neurology (clinical),Radiology, Nuclear Medicine and imaging
Cited by
2 articles.
订阅此论文施引文献