Abstract
AbstractMineral insulated, metal sheathed (MI) Type K and Type N thermocouples are widely used in industry for process monitoring and control. One factor that limits their accuracy is the dramatic decrease in the insulation resistance at temperatures above about 600 °C which results in temperature measurement errors due to electrical shunting. In this work the insulation resistance of a cohort of representative MI thermocouples was characterised at temperatures up to 1160 °C, with simultaneous measurements of the error in indicated temperature by in situ comparison with a reference Type R thermocouple. Intriguingly, there appears to be a systematic relationship between the insulation resistance and the error in the indicated temperature. At a given temperature, as the insulation resistance decreases, there is a corresponding increasingly negative error in the temperature measurement. Although the measurements have a relatively large uncertainty (up to about 1 °C in temperature error and up to about 10 % in insulation resistance measurement), the trend is apparent at all temperatures above 600 °C, which suggests that it is real. Furthermore, the correlation disappears at temperatures below about 600 °C, which is consistent with the well-established diminution of insulation resistance breakdown effects below that temperature. This raises the intriguing possibility of using the as-new MI thermocouple calibration as an indicator of insulation resistance breakdown: large deviations of the electromotive force (emf) in the negative direction could indicate a correspondingly low insulation resistance.
Publisher
Springer Science and Business Media LLC
Reference9 articles.
1. T.J. Quinn, Temperature (Academic Press Ltd., London, 1990), pp. 309–311
2. M.W. Hastings, J.V. Pearce, G. Machin, Electrical resistance breakdown of type N mineral-insulated metal-sheathed thermocouples above 800 °C. Meas. Tech. 55, 941 (2012)
3. J.V. Nicholas, D.R. White, Traceable Temperatures (Wiley, Chichester, 2001), pp. 307–317
4. R.E. Bentley, T.L. Morgan, Ni-based thermocouples in the mineral-insulated metal-sheathed format: thermoelectric instabilities to 1100 °C. J. Phys. E Sci. Instrum. 19, 262–268 (1986)
5. ASTM E608 , E608M-13, Standard Specification for Mineral-Insulated Metal-Sheathed Base Metal Thermocouples (ASTM International, West Conshohocken, 2000)
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献