Abstract
AbstractWe investigated the performance of the qCON index regarding its time delay for sudden changes in the anesthetic level as well as to separate responsiveness from unresponsiveness during loss and return of responsiveness (LOR and ROR). For evaluation of the time delay, we replayed relevant EEG episodes to the qCON to simulate sudden changes between the states (i) awake/sedation, (ii) adequate anesthesia, or (iii) suppression. We also replayed EEG from 40 patients during LOR and ROR to evaluate the qCON’s ability to separate responsiveness from unresponsiveness. The time delays depended on the type of transition. The delays for the important transition between awake/sedation and adequate anesthesia were 21(5) s from awake/sedation to adequate anesthesia and 26(5) s in the other direction. The performance of the qCON to separate responsiveness from unresponsiveness depended on signal quality, the investigation window, i.e. ± 30 s or ± 60 s around LOR/ROR, and the specific transition being tested. AUC was 0.63–0.90 for LOR and 0.61–0.79 for ROR. Time delay and performance during state transitions of the qCON were similar to other monitoring systems such as bispectral index. The better performance of qCON during LOR than ROR probably reflects the sudden change in EEG activity during LOR and the more heterogeneous EEG during ROR.
Publisher
Springer Science and Business Media LLC
Subject
Anesthesiology and Pain Medicine,Critical Care and Intensive Care Medicine,Health Informatics
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献