Accuracy of non-invasive body temperature measurement methods in critically ill patients: a prospective, bicentric, observational study

Author:

Cutuli Salvatore LORCID, ,Osawa Eduardo A,Eyeington Christopher TORCID,Proimos Helena,Canet EmmanuelORCID,Young HelenORCID,Peck Leah,Eastwood Glenn M,Glassford Neil JORCID,Bailey MichaelORCID,Bellomo RinaldoORCID, , , , , , , , , , , , , , , , , ,

Abstract

Objective: The accuracy of different non-invasive body temperature measurement methods in intensive care unit (ICU) patients is uncertain. We aimed to study the accuracy of three commonly used methods. Design: Prospective observational study. Setting: ICUs of two tertiary Australian hospitals. Participants: Critically ill patients admitted to the ICU. Interventions: Invasive (intravascular and intra-urinary bladder catheter) and non-invasive (axillary chemical dot, tympanic infrared, and temporal scanner) body temperature measurements were taken at study inclusion and every 4 hours for the following 72 hours. Main outcome measures: Accuracy of non-invasive body temperature measurement methods was assessed by the Bland–Altman approach, accounting for repeated measurements and significant explanatory variables that were identified by regression analysis. Clinical adequacy was set at limits of agreement (LoA) of 1C compared with core temperature. Results: We studied 50 consecutive critically ill patients who were mainly admitted to the ICU after cardiac surgery. From over 375 observations, invasive core temperature (mostly pulmonary artery catheter) ranged from 33.9C to 39C. On average, the LoA between invasive and non-invasive measurements methods were about 3C. The temporal scanner showed the worst performance in estimating core temperature (bias, 0.66C; LoA, 1.23C, +2.55C), followed by tympanic infrared (bias, 0.44C; LoA, 1.73C, +2.61C) and axillary chemical dot methods (bias, 0.32°C; LoA, 1.64C, +2.28C). No methods achieved clinical adequacy even accounting for significant explanatory variables. Conclusions: The axillary chemical dot, tympanic infrared and temporal scanner methods are inaccurate measures of core temperature in ICU patients. These non-invasive methods appeared unreliable for use in ICU patients.

Publisher

College Of Intensive Care Medicine Of Australia And New Zealand

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3